Main features
- PLC control, colorful touch screen as interface
- Temperature, pressure, time, running status, and error alarm can be dynamic display
- water filling, vacuum, heating, sterilization, exhaust, vacuum drying is fully automatic
- high performance imported vacuum pump is installed, the method of pulse vacuum can draw out
cool air completely
- Post vacuum drying system, humidity residue is less Than 1%
- Built-in steam generator and steam-water separation Device, so high-quality steam can be provided by itself
- Chamber heating film can realize quick preheating and Uniformity temperature.
- Various programs such as instrument, textile, and rubber can be chosen
- BD test program and leak test program are preset; Leakage rate is less than 0.13kpa/min and it can pass PCD test
- Buzzer will alarm for cycle end
- it has maintenance hint function so that user can maintain autoclave easily
- Printer is installed, it can print date, and Time and process, 15-99 seconds can be set for print interval
Safety Devices
Auto protection device against over temperature: this system cuts off
Power automatically if the designed temperature is surpassed.
Door safety interlock: the door cannot be opened until there is no
Pressure in the chamber and power for heating cannot be switched on Until the door is firmly closed.
Safety valve of auto release against overpressure: it opens when the
Designed pressure is surpassed, the program will be stopped if water level lower than the low limit,
Alarm with sound and light.
Electronic circuits safety device: direct current control loop and has
Protection against overload and voltage surpassed, alternating current
Main loop has short-circuited protection device.
Main technical parameters
- ·Designed pressure: ﹣0.1/0.28Mpa ·Material of chamber: US 304 stainless steel
- ·Sterilization temperature:105℃—136℃ ·Maximum working pressure:0.23Mpa
- ·Temperature degree of uniformity: ≤±0.5℃ ·vacuum limit: -0.080Mpa
- ·Accuracy of pressure display:0.1Kpa ·Precision of temperature displayed:0.1℃